The heating of this incubator is optimally tuned for both natural convection and forced air circulation; the fan can also be switched off completely, and valuable chamber loads for research, pharmaceutics, medicine and food chemistry are warmed up very carefully.


Feature

Temperature

Working temperature range
: min. 10°C above ambient up to +80°C.
Setting accuracy temperature
: 0.1°C.
Temperature
: 1 Pt100 sensor DIN class A in 4-wire-circuit.

 

 

 

Control technology

Language setting
: German, English, Spanish, French, Polish, Czech, Hungarian
ControlCOCKPIT
: adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-       colour display
Timer
: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT
: the process time does not start until the set temperature is reached
Calibration
: three freely selectable temperature values
adjustable parameters
: temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time     zones, summertime/wintertime

 

 

Ventilation

Fan
: forced air circulation by quiet air turbine, adjustable in 10 % steps.
Fresh air admixture
: adjustment of pre-heated fresh air admixture by air flap control in 10 %         steps.
Vent
: vent connection with restrictor flap.

 

 

Safety

Temperature control
: adjustable electronic overtemperature monitor and mechanical temperature   limiter TB, protection class 1 according to DIN 12880 to switch off the             heating approx. 20°C above nominal temperature.
Autodiagnostic system
: for fault analysis.