Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.



Working-temperature range
5 °C above ambient temperature up to +50 °C 
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
Setting temperature range
+18 to +50 °C
Setting accuracy temperature
0.1 °C
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value


Setting range active humidity control
40 to 97 % rh and rh-Off
Setting accuracy humidity
0.5 % rh
Humidity control
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %

Control of standard components

CO2 control
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
Adjustment range CO2
0 to 20 % CO2
Variation in time CO2
+/- 0.2 % CO2
Adjustment range O2
1 to 20 % O2
Setting accuracy O2
0.1 % O2

Control technology

TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting
German, English, Spanish, French, Polish, Czech, Hungarian
Function SetpointWAIT
the process time does not start until the set temperature is reached
adjustable parameters
temperature (Celsius or Fahrenheit), fan speed, CO2, programme time, time zones, summertime/wintertime


Ethernet LAN, USB
programme stored in case of power failure
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port


Autodiagnostic system
integral fault diagnostics for temperature, CO2 and humidity limit control
visual and acoustic


Heating concept

6 sides
large-area multi-function heating system on four sides with additional door and back heating to avoid condensation

Standard equipment

Works calibation certificate °C & CO2
incl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard units
Works calibration cerfificate °C, CO2 & rH
incl. works calibration certificate for +37°C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for units with active humidity control
Works calibration certificate °C, CO2, rH & O2
incl. works calibration certificate for +37°C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option T6); Standard equipment for units with O2 control
1 pc stainless steel shelf, perforated
1 pc stainless steel water dish
fully insulated stainless steel door with 2-point locking (compression door lock)
inner glass door with opening (8 mm Ø) to take gas sample

Stainless steel interior

material 1.4301 (ASTM 304), corrosion resistant
56 l
Dimensions W x H x D in mm
w(A) x h(B) x d(C): 400 x 425 x 330 mm
Max. number of grids/shelves
Max. loading of chamber
75 kg
Max. loading per grid/shelf
15 kg

Textured stainless steel casing

w(D) x h(E) x d(F): 559 x 791 x 521 mm
rear zinc-plated steel

Electrical data

Voltage Electrical load
230 V, 50/60 Hz 
approx. 1000 W
Voltage Electrical load
115 V, 50/60 Hz 
approx. 1000 W