Home  |   Contact Us
Home   »   Produk   »   Water Quality Equipments

Eutech PD 450

 

pH / mV / Dissolved Oxygen / Temperature Aordable optical RDO technology reduces maintenance time and cost


Feature

  • Maintenance free optical probe with no electrolyte eliminates hassle of replacing delicate membranes and salt build up from refilling solutions
  • Switch between pH, mV, % saturation and mg/L (ppm) dissolved oxygen – With the press of a button
  • Durable waterproof and dustproof design with IP67 rating
  • Large, backlit graphic display provides measurement with temperature
  • Calibrate up to 5 points automatically with USA or NIST buffer groups (up to 6 points with DIN). New manual option allows calibration with up to 5 custom buffer standards
  • Auto buer recognition – Automatically identifies the correct pH buffer for rapid calibration
  • Calibration due alarm – Set a visible reminder when it is time to recalibrate the meter
  • Selectable ºC or ºF – Choose units to best fit your application
  • Ready indicator, manual or automatic measurement hold feature and optional automatic shutoff function
  • Store up to 500 data sets – Easy USB or RS232 output with optional data cables
  • Built-in real-time clock stamps stored data and calibration data with time and date –Meets standards for GLP (Good Laboratory Practices)

Spesification

PH Range : -2.00 to 16.00 pH
  Resolution : 0.01 pH
  Accuracy : ±0.01
  Calibration : Up to 6 points: USA, NIST, DIN or custom
MV Range : ±2000 mV
  Resolution : 0.1 mV ±999.9 mV; 1 mV beyond
  Accuracy : ±0.2 mV +2 LSD or +0.5 % of reading; whichever is greater
  Calibration : ±200 mV
Dissolved Oxygen Range : 0 to 200 %; 0 to 20 ppm/mg/L
  Resolution : 0.10 %; 0.1 ppm/mg/L
  Accuracy : ±2 %; ±0.1 ppm or mg/L up to 8; / ±0.2 ppm or mg/L above 8
  Calibration : Zero and saturation
     
Temperature Range : -10.0 to 110.0 ºC (14.0 to 230.0 ºF)
  Resolution : 0.1 ºC/ºF
  Accuracy : ±0.5 ºC; ±0.9 ºF
  Calibration : Offset in 0.1 increments, ±10 ºC or ±18 ºF
Display : Custom LCD with backlight  
Memory : Stores up to 500 data sets  
Output : USB or RS232  
Real-Time Clock : Time and date stamp on calibration and stored data  
Temperature
Compensation
: manual  
Power : Two AA batteries (included), >300 hours (without backlight);  
  AC adapter (optional)