Home  |   Contact Us
Home   »   Produk   »   Water Quality Equipments

Portable pH/ORP/ION meter D-73

 

  

      


Feature

  • The built-in data memory (1000 items)
  • Connecting to a computer allows measurement data to be collected.
  • Output to a GLP/GMP-compatible printer is possible.
  • Automatic calibration and calibration interval alarm function
  • Usable with both 5-point calibration and USA/NIST standard solutions.
  • Interval measurement function
  • Sample ID No. setting function
  • Clock function and auto power-off function
  • Usable with AAA alkaline batteries, Ni-MH batteries, or adapter.
  • IP67-compliant waterproof and dustproof structure*.

Spesification

MODEL

D-73

pH

Measuring principle

Glass electrode method

Measuring range

pH 0.00~14.00

Display range

-2.00~16.00 ※Flashing in the case of outside measurement range.

Resolution

0.01 pH

Repeatability

±0.01 pH±1digit

Auto calibration (5 points)

Calibration record

Standard solution Auto-detect

USA/NIST selectable

Calibration interval alarm

mV(ORP)

Measuring range (Display range)

-2000~2000 mV
※Flashing in the case of outside measurement range.

Resolution

1 mV

Repeatability

±1 mV±1digit

Absolute. relative selectable

Temperature

Measuring range (Display)

0.0℃~100.0℃(-30℃~130℃) 
※Flashing in the case of outside measurement range.

Resolution

0.1℃

Repeatability

±0.1℃±1digit

Calibration function

Ion

Measuring principle

Ion electrode method

Measuring range (Display range)

0.00 μg/L~999 g/L

Resolution

3-digit valid numbers

Repeatability

±0.5% F.S.±1digit

Auto range

2 cH measurement

g/L、mol/L selectable

5 points calibration

Calibration record

Display

Custom LCD with backlight

Function

PC connectivity*1

Printer connectivity(GLP/GMP)

Temperature compensation(Auto/manual)

Auto Hold function

Data memory number

1000

Interval memory

ID input

Clock function

Auto power off

Battery Level Indicator

Dustproof and waterproof standard

IP67

Operating ambient temperature/ humidity

0℃ to 45℃ , 80% or less in relative humidity (no condensation)

Power

LR03/AAA alkaline batteries or AAA Ni-H rechargeable batteries × 2
AC adapter 100 V to 240 V 50/60 Hz (option)

Current consumption

Less than 2 mA

Battery life*2

Approx. 500 hours

Dimension

Approx. 67 (80) × 28 (42) × 170 mm (The figures in parentheses are maximum thicknesses.)

Weight (without batteries and electrode)

Approx. 285 g

*1 RS-232C cable(3014030151) and software is required. Software can be download by web registration. If you need to connect to the USB, the adapter(RS232C⇔ USB) commercially available is required. Please buy one that be suitable for the specifications of the PC (OS · USB Specification, etc.). ※ HORIBA will not guarantee the adapter operation.
*2 Battery life will be shorter when using optional accessories and backlight LCD is activated.