Home  |   Contact Us
Home   »   Produk   »   Oven Laboratorium

Oil Content Analyzer OCMA 500

 

Compact Oil Content Analyzer new OCMA-500 series  focuses  on enhanced operability and useful functions  enabling  anyone  to handle with no hassle. Offering two models, OCMA-500 series is utilized for various purposes including wastewater monitoring, environmental conservation, and quality control.


Feature

  • Allows measurement of any oil with low boiling point (NDIR method)
  • Easy and Speedy measurement
  • Color graphic display with multi-language (English/Russian/Japanese)
  • USB data output port
  • Unit conversion function (mg/L, mg/kg, mg/g, mg/PC)

Spesification

Measurement principle:

Solvent extraction, NDIR
analysis (Infrared spectroscopy)

Measurement range:

0mg/l ~ 200mg/l dynamic ranging

Resolution

0 to 99.9 mg/L; 0.1 mg/L
100 to 200 mg/L; 1 mg/L

Repeatability

0 to 9.9 mg/L; ± 0.2 mg/L ± 1 digit
10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit
100 to 200 mg/L; ± 4 mg/L ± 1 digit
*For standard liquids 

Display Method

3.5 inches, 320×240 dots
Backlight Colour graphic LCD

Calibration Method

Zero, Span Calibration

Extraction Solvent

S-316 solvent

Amount of test sample required

Sample water : Solvent = 2 : 1 

Amount of extraction solvent required

8 mL

Extraction Method

Built-in extractor

Ambient Operationg Temperature

0 to 40 ℃ (non condensation)

Power supply

AC 100V to 240V ±10%, 50/60 Hz

Power consumption

AC 100V:Approx.60VA
AC 240V: Approx.90VA

External Dimensions

200(W)×313(D)×342(H) mm

Mass

Approx. 7 kg

External Output

Output to an USB memory stick

Functions

 

300-item data memory
Stabilized measurement value display
Self error determination
Clock